A flue gas analyser is only as good as the routine you run it through. Skip the warm-up or the fresh-air zero and the readings can mislead you — and on a combustion check, a misleading reading is a safety problem. This guide walks through using an electronic combustion gas analyser (ECGA/FGA) properly and reading what it shows. It's study material; only a Gas Safe registered engineer may carry out the work.

Standard. Combustion analysis is carried out to BS 7967 (current edition 2015), using an analyser conforming to BS EN 50379-3. Always work to the appliance manufacturer's instructions, which take precedence and tell you the sampling point and the pass figures.

Before you start: prepare the instrument

  1. Zero in fresh air. Switch on and let the analyser zero its sensors in clean outdoor air, away from any flue, vehicle exhaust or your own breath. This sets the baseline everything else is measured from.
  2. Let it warm up. The sensors need a short stabilisation period before they read accurately — don't rush to the flue.
  3. Check the consumables. Empty the water trap and check the particulate filter is clean. Moisture or soot reaching the sensors gives false readings and shortens their life.
  4. Confirm it's in calibration. A reading from an out-of-calibration analyser isn't valid (more on this in the care & calibration guide).

Sampling the products

Insert the probe at the appliance's purpose-designed sampling point — not a random hole — to the depth the manufacturer specifies, so you sample the true combustion products and not diluted air. Run the appliance to the test condition (often maximum rate) and let the readings stabilise before recording. Take readings at the rate(s) the manufacturer asks for.

What each reading means

The maths the analyser does for you. Complete combustion of methane is CH₄ + 2O₂ → CO₂ + 2H₂O — carbon dioxide and water, no CO. Any CO at all means the reaction was incomplete. The analyser combines CO and CO₂ into the CO/CO₂ ratio — the single most important output, covered in its own guide.

After the test

When you're finished, run the analyser in fresh air again to purge the sensors of flue gas before switching off — it protects the sensors and leaves the instrument ready for next time. Record your readings (for a boiler, on the Benchmark log) and cap the test point.

  1. Zero in fresh air, warm up, empty the water trap, check the filter, confirm calibration.
  2. Sample at the manufacturer's purpose-designed point, to the right depth, at the right rate.
  3. Let readings stabilise before recording.
  4. CO = safety (ppm); CO₂ = reference (%); O₂ = excess air; temperature = efficiency.
  5. Any CO means incomplete combustion — the CO/CO₂ ratio is the key output.
  6. Purge in fresh air before switching off, then record and cap the test point.
  7. Manufacturer's instructions set the sampling point and the pass figures.

10-Question Mock Test

Click an option to see whether you got it right. Explanations appear instantly — no submitting at the end.

Your score: 0 / 10
Question 1 of 10
Where must an analyser be zeroed before sampling?
Question 2 of 10
Why let the analyser warm up before use?
Question 3 of 10
What is the water trap on an analyser for?
Question 4 of 10
Where should the probe sample?
Question 5 of 10
When should you record the readings?
Question 6 of 10
A CO reading is given in:
Question 7 of 10
The oxygen (O₂) reading mainly tells you about:
Question 8 of 10
Complete combustion of methane produces:
Question 9 of 10
What should you do with the analyser when the test is finished?
Question 10 of 10
For the sampling point and pass figures, what takes precedence?

Good readings start with good routine. Make it second nature.

PlumbMate drills the CPA1 method — zero, warm-up, sample, purge — with quizzes and spaced repetition mapped to the gas ACS tickets.

🔒 PlumbMate Gas — coming soon

Full ACS revision — CCN1, CPA1, CKR1, HTR1 & CENWAT · £29.99/year · Launching soon

Prefer to browse first? Back to the Gas Blog →